Mahr Millimar C 1200 Operating Instructions Manual page 38

Compact length measuring instrument
Hide thumbs Also See for Millimar C 1200:
Table of Contents

Advertisement

Enter value for Calibration Max
(e.g. +1.000 mm). Press u
select the required position and
press q p to enter the required
value.
ü
Press
to confirm or Ó
the selection
Move the Probe to the Min
position (e.g. measuring plate)
=>
The following appears on the
display:
Set probe to Calibration Min and
accept
ü
Press
to confirm
Move the probe to the Max
position (e.g. 2.000 mm gauge
block)
=>
The following appears on the
display:
Set probe to Calibration Max and
accept
ü
Press
to confirm
=>
Probe calibration according to
the customer's specification is
complete and activated.
Press the ON / OFF key to exit the
input menu.
38
to
to cancel
Measurement
Display
Setup
Customer calibration
+ 0 0
1
. 0 0 0
Calibration Min
Nominal value
-1.000
-0
ü
u
Ó
Measurement
Display
Setup
Customer calibration
Calibration
Min
and accept
Calibration Min
Nominal value
-1.000
-0
ü
3
Ó
Display
Setup
Measurement
Customer calibration
Calibration
Max
and accept
Calibration Min
Nominal value
-1.000
-0
ü
3
Ó
Measurement
Display
Setup
Absolute mode
Customer calibration
Factory calibration
Factory settings
Keyboard-/ menu interlock
Info
ü
Ó
Mahr GmbH • Millimar C1200
System
Calibration Max
1.000
q
p
System
Calibration Max
1.000
q
p
System
Calibration Max
1.000
q
p
System
ü
q
p

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Millimar C 1200 and is the answer not in the manual?

Questions and answers

Table of Contents