Preventive Maintenance - Thermo Scientific 370 Operating And Maintenance Manual

370 series steri-cycle co2 incubator direct heat with sterilization cycle
Hide thumbs Also See for 370:
Table of Contents

Advertisement

Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to
keep your unit functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For
maximum performance and efficiency, it is recommended the unit be checked and calibrated periodically by a qualified service
technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the instruction manual for
further details.
We have qualified service technicians, using NIST traceable instruments, available in many areas. For more information on
Preventive Maintenance or Extended Warranties, please contact us at the number listed below.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Tips for all incubators:
Do NOT use bleach or any disinfectant that has high chloros
Use sterile, distilled or demineralized water.
USA and Canada 888-213-1790 • Telefax: 740-373-4189 •
Refer to Manual
Action
Section
--
Check CO2 tank levels.
--
Inspect door latch, hinges and door gasket seal.
1
Check water level in the humidity pan, ½" from top.
2
Verify and document CO2, humidity and temperature calibration, as
applicable (See Calibration).
Disinfect the interior of the incubator (See Routine Maintenance).
1
Replace HEPA, access port filter, air sample filter, and CO2 filters*, if
applicable (or as needed).
5
Perform sterilization cycle as needed.
Thermo Fisher Scientific
401 Millcreek Road, Box 649 • Marietta, Ohio 45750 USA • 740-373-4763
Preventive Maintenance for Model 370/380 Series Incubators
Avoid spraying cleaner on the CO
Do not use powdered gloves for tissue cultures.
http://www.services.controlenv@thermo.com
Daily
Weekly
Monthly

Preventive Maintenance

sensor.
2
3 to 6
Yearly
Months
Steri-Cycle CO
Incubator
2