Installing the HEPA
Filter (continued)

Leveling the Unit

Connecting the Unit
to Electrical Power
Thermo Fisher Scientific
1. Remove the filter from the shipping box.
2. Remove the plastic coating from the filter, using caution not to touch
the filter media.
3. Install the filter as shown in Figure 1-13. Refer to Section 5 for HEPA
filter maintenance.
Figure 1-13. Air Sample Filter Installation
Check for level by placing a bubble-style level on one of the shelves. Turn
the hex nut on the leveler counterclockwise to lengthen the leg or
clockwise to shorten it. Level the unit front-to-back and left-to-right.
See the serial tag on the side of the unit for electrical specifications or refer
to the electrical schematics at the end of this manual.
Caution Serial tag amp rating is based on amperage draw during
sterilization cycle. Normal operating amperage is much less. Ensure that
electrical circuit will handle amp draw of sterilization cycle.
Connect the incubator to a grounded, dedicated circuit. The power cord
connector is the mains disconnect device for the incubator. Position the
incubator so the unit can be easily disconnected.
Section 1
Installation and Start-Up
Steri-Cycle CO
Incubator
2
1-9