16. ANALOG 4-20 mA OUTPUT

16.1. Terminal Assignment

Terminal No.
Symbol
8
A+
9
A-
10
SHLD
When the optional analog 4-20 mA output is not installed, the terminals No.8 and No.9 are vacant.
Check the 07 column of the option list on the side of the indicator casing.

16.2. Connection Diagram

Applicable solderless terminal: R1.25-3
Applicable wire size: 0.3 to 0.75 mm
External load resistance: 0 to 600 
AD-4401A
Description
Analog current output +
Analog current output -
Frame ground
Shield
Analog current +
Analog current -
2
. Use a 2-wire shielded twisted pair cable.