●Tare
The current gross value is saved as a tare value and the net value is set to zero.
Operation: TARE key, Control input
Setting: Tare and zero setting at unstable weight (CALF-10) / Tare at negative gross value (CALF-11)
●Tare clear
The tare value is cleared.
Operation: Function key (FncF-02), Control input
●Gross/Net value selection
The weight value display method is selected, gross value or net value. The GROSS status turns ON when
the gross value is displayed. The NET status turns ON when the net value is displayed.
Operation: NET/GROSS key, Control input
●Zero
Center of zero is detected and the ZERO status turns ON when the gross value is within 1/4 of the minimum
division.
●Stable
Stability is detected when the weight value is maintained within the stability detection width (CALF-09) during
the stability detection time (CALF-08). The STABLE (MD) status turns OFF. The control output that is set to
"Stable" turns ON.
●Hold
The weight value is temporarily locked / released. The HOLD status turns ON while the weight value is
locked.
Operation: Function key (FncF-02), Control input
Select a hold function (FncF-08) from the following:
- Normal hold: The value when the holding operation is performed is locked.
- Peak hold: The maximum value after the holding operation is locked.
Setting: Comparison at hold (FncF-09)
●Manual Print
Data is output from the standard serial output or options (BCD, RS-232C, RS-422/485).
Operation: Function key (FncF-02), Control input
●Accumulation
The net value is accumulated on the total weight and the number of accumulations is updated.
Operation: Function key (FncF-02), Control input