Technical Specifications - biochrom Novaspec III Plus Manual

Table of Contents

Advertisement

Technical Specifications

Wavelength range
Monochromator
Wavelength calibration
Spectral bandwidth
Wavelength accuracy
Wavelength reproducibility
Light sources
Detector
Photometric range
Photometric linearity
Photometric reproducibility
Stray Light
Stability
Noise
Digital output
Data Export
Method Storage
Graphical Display
Sample ID
Languages
Dimensions
Weight
Power input
325nm to 1100nm
Flat grating
Automatic upon switch on
<7nm
± 2nm
± 1nm
Pulsed Tungsten halogen
CMOS array
- 0.300 to 2.500A, 0.3 to 199%T
± 2.0 % or ± 0.010A to 1.000A at 546nm,
whichever is the greater
< 0.002 A at 0A and 500nm
< 1%T 340nm
± 0.005A/h at 0A and 546nm after warm-up
± 0.002A near 0A and ± 0.020A near 2A at
600nm
USB memory stick data storage
USB PC connection with PVC software
PC via USB cable
USB memory stick
90 with PIN number protection
Yes, zoom and track function
Yes
English, German, French, Spanish, Italian,
Japanese, Chinese
120 x 230 x 320 mm
1.55 kg
18 Vdc at max 18VA from a supplied 100 to
240V~, 50/60 Hz Mains Power Adapter
6 |
P a g e

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Novaspec III Plus and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF