Concentration with a Standard

Single point calibration curve with known standard and need to calculate a factor.
Step 1
Press the Options Key
Step 2
Press 1 to set parameters.
Step 3
If not already in Conc. Standard mode use left or
right arrow keys to Conc Standard mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
Using the keypad, enter the concentration.
Press the down arrow key.
Step 6
The units can be manually written in using the
keypad, or press
arrow keys to select the units (µg/ml, µg/µl, pmol/µl,
mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm,
ppb, conc or none).
This screen also allows the number of displayed
decimal points (DP) to be selected, from 0 to 2.
Note that the result will always be fixed to 5
significant figures regardless of how many decimal
points are selected (so 98768.2 will display as
98768 even with 1 decimal point selected).
.