72
Temperature derating for FMP52
[°C] ([°F]) T a
T a
+80 (+176)
T p
GT19: -36 (-33)
GT18: -37 (-35)
GT20: -38 (-36)
[°C] ([°F]) T a
GT18/20: +79 (+174)
GT19: +74 (+165)
GT19: -36 (-33)
GT18: -37 (-35)
GT20: -38 (-36)
[°C] ([°F]) T a
+81 (+178)
GT18: -37 (-35)
GT20: -38 (-36)
[°C] ([°F]) T a
+79 (+174)
GT18: -37 (-35)
GT20: -38 (-36)
GT18 = stainless steel housing
GT19 = plastic housing
GT20 = aluminum housing
1)
For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the
switch output. (G
1
: switch output not connected; G2: switch output connected).
2)
For saturated steam applications the process temperature should not exceed 150 °C (302 °F). For higher
process temperatures use FMP54.
A:
4 20 mA HART
–
-40 (-40)
-50
-40
(-40)
(-58)
C:
4 20 mA HART
–
4 20 mA
–
K:
90–253 VAC
L:
10.4–48 VDC
-40 (-40)
-50
-40
(-58)
(-40)
1
G :
PROFIBUS PA
FOUNDATION Fieldbus
-40 (-40)
-50
-40
(-58)
(-40)
2
G :
PROFIBUS PA
FOUNDATION Fieldbus
Switch output
-40 (-40)
-50
-40
(-58)
(-40)
A = 1 current output
C = 2 current outputs
1)
G
1
, G
2
= PROFIBUS PA
K, L = 4-wire
Levelflex FMP51, FMP52, FMP54
+82
+200
(+180)
(+392)
+79
(+174)
+74
+200
(+165)
(+392)
+81
+200
(+178)
(+392)
+79
+200
(+174)
(+392)
T
= ambient temperature
a
T
= temperature at the process connection
p
Endress+Hauser
GT20: +60 (+140)
GT18: +53 (+127)
GT19: +38 (+100)
T p
[°C]
([°F])
GT20: +56 (+133)
GT18: +51 (+124)
GT19: 38 (+100)
T p
[°C]
([°F])
GT20: +59 (+138)
GT18: +53 (+127)
T p
[°C]
([°F])
GT20: +56 (+133)
GT18: +51 (+124)
T p
[°C]
([°F])
A0013633
2)
Need help?
Do you have a question about the Levelflex FMP51 and is the answer not in the manual?