Levelflex FMP51, FMP52, FMP54
Endress+Hauser
Temperature derating for FMP51 with flange
[°C] ([°F]) T a
T a
+80 (+176)
T p
-40 (-40)
[°C] ([°F]) T a
GT18/20: +79 (+174)
GT19: +74 (+165)
-40 (-40)
[°C] ([°F]) T a
+81 (+178)
-40 (-40)
[°C] ([°F]) T a
+79 (+174)
-40 (-40)
GT18 = stainless steel housing
GT19 = plastic housing
GT20 = aluminum housing
1)
For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the
1
switch output. (G
: switch output not connected; G2: switch output connected).
2)
For saturated steam applications the process temperature should not exceed 150 °C (302 °F). For higher
process temperatures use FMP54.
A:
4 20 mA HART
–
-40
+82
(-40)
(+180)
+79
(+174)
C:
4 20 mA HART
–
4 20 mA
–
K:
90–253 VAC
L:
10.4–48 VDC
-40
+74
(-40)
(+165)
1
G :
PROFIBUS PA
FOUNDATION Fieldbus
-40
+81
(-40)
(+178)
2
G :
PROFIBUS PA
FOUNDATION Fieldbus
Switch output
-40
+79
(-40)
(+174)
A = 1 current output
C = 2 current outputs
1)
G
1
, G
2
= PROFIBUS PA
K, L = 4-wire
GT20: +59 (+138)
GT18: +53 (+127)
GT19: +33 (+91)
T p
[°C]
+200
([°F])
(+392)
GT20: +55 (+131)
GT18: +51 (+124)
GT19: +33 (+91)
T p
[°C]
+200
([°F])
(+392)
GT20: +58 (+136)
GT18: +53 (+127)
T p
[°C]
+200
([°F])
(+392)
GT20: +55 (+130)
GT18: +51 (+123)
T p
[°C]
+200
([°F])
(+392)
T
= ambient temperature
a
T
= temperature at the process connection
p
A0013689
2)
71
Need help?