4. Filter cubes for fluorescence observation (for the LV-UEPI2 only)
The LV-UEPI2 accommodates two filter cubes for epi-fluorescence microscopy.
1 Verify that the illumination shutter is closed and
the power supplies to the microscope and light
source are off.
2 Remove the cover from the left side of the
illuminator.
3 Turn the microscopy selection knob so that the
position indication "FL1" or "FL2" on the turret
in the illuminator faces the opening.
4 Insert the desired filter cube into the dovetail of the
turret and push it in to the clickstop position.
Make sure that the filter cube has its excitation
filter facing out.
5 Now that the filter cube is installed in the position
FL1 or FL2, refit the cover.
6 Check the stickers of excitation method supplied
with the illuminator and find the one that
corresponds to the filter cube just installed. Affix it
to the position FL1 or FL2 on the microscopy
selection knob.
If there is no sticker corresponding to the
excitation method of the filter cube, write the
excitation method in a blank sticker and affix it.
52
Cover
UV -2A
EX 330-3
80
DM 400
BA 420
Filter cube
FL1
Sticker to be affixed to the
microscopy selection knob
UV-2A
UV-2B
UV-1A
V-2A
EX 330-380
EX 330-380
EX 365/10
EX 380-420
DM 400
DM 400
DM 400
DM 430
BA 420
BA 435
BA 400
BA 450
BV-1A
B-3A
B-2A
B-1A
EX 435/10
EX 420-490
EX 450-490
EX 470-490
DM 455
DM 505
DM 505
DM 505
BA 470
BA 520
BA 520
BA 520
G-2A
G-2B
G-1B
EX 510-560
EX 510-560
EX 546/10
DM 575
DM 575
DM 575
BA 590
BA 610
BA 590
Stickers of excitation method
UV-2A
EX 330-380
DM 400
BA 420
BV-2A
EX 400-440
DM 455
BA 470
B-1E
EX 470-490
DM 505
BA 520-560