Transmitter 9135 - pH/redox measurement
ALARM 3
Sys
MODE :
ACCEPT : Manu
RELAY : NO
Select
ALARM 4
Timer
MODE :
INTERV : 2440 mn
IMPUL. Nb : 1
Ton : 005 s
Toff : 005 s
TmA : 00 mn
Select
Page 46

ALARM 3 (Alarm system)

Choice of a limit on the measure or alarm
system with alarm 3.
MODE :
Manual or automatic acquit possible with the
alarm system.
ACCEPT : •
Choice between relays normally open or closed.
RELAY : •

ALARM 4 (Timer)

Choice between a limit on the measure or a
timer for the probe cleaning with alarm 4.
MODE :
INTERV
: interval between 2 cleaning cycles
in minutes.
IMPUL. Nb : number of pulses during a cleaning
cycle.
Ton
: time when relay is activated , in
secondes.
Toff
: time when relay is desactivated, in
secondes.
TmA
: analog output hold time for the
analogue outputs in minutes.