(5) Press the [ENTER] key to start the Setup3 (RTD temp.) operation.
(6) The
"0 ohm :
[START]" message is displayed at the location shown above.
(7) Connect the reference 0 Ω resistor to the input terminals of channel 20.
(8) Press the [START] key and then wait until the
Replace the reference resistor to correspond to one of the displayed resistors in the following table and
then press the [START] key.
Message
0 ohm
350 ohm
2500 ohm
(9) Repeat step (8) by sequentially to 350 Ω and 2500 Ω.
(10) The "EEPROM : [ENTER]" message appears when the 2500 ohm setup was completed.
(11) Press the [ENTER] key to begin writing the new Setup data to the EEPROM.
(12) After the setup parameters have been registered to the EEPROM, the display closes and returns to
normal mode.