Auxiliary Wall Outlets; Fore Vacuum Pump; Quality Of Power - Thermo Scientific iCAP RQ ICP-MS Pre-Installation Requirements Manual

Hide thumbs Also See for iCAP RQ ICP-MS:
Table of Contents

Advertisement

Line Power
Electrical Power Requirements

Auxiliary Wall Outlets

Fore Vacuum Pump

Quality of Power

Table 4-3.
Immunity compliance of iCAP RQ ICP-MS
Description
Immunity to Electrical Fast Transient/Burst
Immunity to Electrical Slow Transient/Surge
Immunity to Conducted RF Voltage
Immunity to Voltage Dips, Short Interruptions and Voltage
Variation
4-4
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Additional single-phase outlets are needed for additional parts such as
fore vacuum pump, computer and monitor. Thermo Fisher Scientific
recommends at least five spare outlets in the near vicinity of the left side
of the system and five close to the workbench space within the your
laboratory.
The components of the data system (computer, monitor) require wall
outlets at a nominal voltage of 200-240 V AC, 50/60 Hz, also see the
manufacturer's manual for more information.
The fore vacuum pump requires a separately fused wall outlet due to the
high current of the pump. This wall outlet must be fused with 15 A or
16 A, tripping characteristic 'C' or 'D'.
For information on the power outlet requirements, also see the
manufacturer's manuals.
The fore vacuum pump requires a dedicated power outlet that complies
with the specifications as indicated on the rating plate of the fore
vacuum pump, see
Table
Table 4-2.
Power requirements for fore vacuum pump
Voltage
200-240 V ±10%
200-240 V ±10%
The iCAP RQ instrument complies with the requirements listed in
Table
4-3. See
Chapter 6: "General Pre-Installation Information"
details on the required quality of power.
4-2.
Frequency
50 Hz
60 Hz
Requirement
EN 61326-1, EN 61000-4-4
EN 61326-1, EN 61000-4-5
EN 61326-1, EN 61000-4-6
EN 61326-1, EN 61000-4-11
Power
1.6 kW
1.9 kW
for
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents