Restoring Factory Protocols

If necessary, you can choose to restore the unit's original (default) waveform parameters when it was shipped to you.
NOTE: This procedure will erase all user-defined protocols.
To restore the unit's original waveform parameters, do the following:
1.
On the Home screen, press the Clinical Library button.
The Clinical Library window displays.
2.
Press the Up Arrow or Down Arrow buttons to highlight Restore Factory
Protocols.
3.
Press the Enter button to accept the highlighted selection.
The unit displays the message "Restoring Protocols Please wait. "
The user-defined protocols are erased and the waveforms are restored to the original
parameters. Then the Restore Factory Protocols Confirmation window displays.
4.
Press any button on the Operator Interface.
You are returned to the Clinical Library window.