Memory

To activate the Memory feature, press
the A and ON buttons together at power-
up (Fig. 16). MEM will be displayed.
The Memory feature allows the user to
log into memory 1 to 99 measured values
(including OL). Measured values can
now be logged for both resistance (Ω)
and amps (A) by pushing the MEM button
during measurement and during "HOLD".
A long beep indicates the measurement
was stored. After the 99th value is entered (MEM 99) will blink indicating
memory is now full. Measurements may still be made but no longer
logged. Stored values are retained when the instrument is turned off.
Pressing the A and ON buttons again at power-up will de-activate the
Memory feature.

Memory Recall

AEMC
Model 3730
GROUND TESTER
I N S T R U M E N T S
MR
MEM
AL
A
Figure 17

Clear Memory

To clear the memory, push ON & MEM together at power-up and hold
them down. After a three second interval the "clear" message will be
displayed (Fig. 18) along with four repetitive beeps and a fifth long tone.
The memory is now cleared (Fig. 19).
Once cleared, releasing the ON & MEM buttons returns the instrument
to Resistance measurement mode.
Ground Resistance Tester Models 3710/3730
AEMC
I N S T R U M E N T S
MEM
To recall memory values, push ON and
MEM together at power-up. MR will be dis-
played with the memory number (Fig. 17).
Memorized value will be displayed. To
select stored values use the cursor arrows
(
down,
up). For rapid access of
memory recall values, press the cursor
ON
OFF
push-button continuously.
Turn the instrument off to exit the Memory
Recall mode.
Model 3730
GROUND TESTER
MEM
ON
AL
A
OFF
Figure 16
19