Calibration Check Loop
The multiple calibration loop is now replaced by a single point 25 Ω loop
(Fig. 33). It is provided to perform a quick check of the instruments.
Check the clamp-on ground resistance tester by clamping around the
loop; the 3710 or 3730 should read between 24.2 Ω and 25.8 Ω.
* Value is for ambient temperature between 68° and 78°F. Value may differ a
few counts under or above this temperature.
®
I N S T R U M E N T S
CALIBRATION LOOP 25Ω
Figure 33
32
Ground Resistance Tester Models 3710/3730