13.2.2 Lamp change

+ Disconnect equipment from power supply.
+ Only change the lamp when the equipment has cooled down.
1. Remove protective glass guard, see chap.
13.2.1
2. Remove plug connector [3]. Remove cable from
terminals [2] with care.
3. Loosen lamp [1] from the clips [4] on both sides.
4. Install new lamp in reverse order.
Avoid contact with the halogen lamp to maintain service life.
Position plug connector in accordance with illustration.
Fit protective glass guard, see chap. 13.2.1