Change Battery - KERN DLB-A01 Operating Instruction

Temperature calibrating set
Table of Contents

Advertisement

Available languages

Available languages

3 Change battery

If the battery voltage is not more sufficient, the battery symbol
play.
+ Switch off appliance and remove cable from temperature sensor plate.
+ Open the lid of the battery compartment loosening the screw.
+ Replace battery.
Ensure correct polarisation.
+ Close again the lid.
4 Adjust/calibrate the temperature value of the KERN DLB-A
4.1 Preparation
We recommend sometimes to check the temperature value of the device using the op-
tional temperature calibrating set DLB-A01. Before you do this, allow the device to cool
down for at least 3 hours after the last heating phase.
Remove the separate component parts „sample dish" according to their sequence
12
1.
2.
3.
4.
will appear in the dis-
DLB_A01-BA-def-1110