14.1 Error messages

Error message Explication
ERR.001
ERR.002
Hardware error
ERR.005
Memory error
Heated cover is open for more
ERR.100
than 1 minute during
measuring.
ERR.101
Failure „Temperature sensor"
ERR.102
Heated cover not closed
ERR.110
properly
Measuring start 30 minutes
TIM.oUT
after resetting to zero
ERR.121
ERR.122
Failure „Heating"
ERR.123
ERR.124
Measuring lasts too long
ERR.200
Failure „Current supply"
ERR.201
Internal error
ERR.202
Failure „Electrical voltage"
DBS-BA-e-1311
Remedy
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Press ESC to abort measuring.
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Press ESC to abort measuring.
Press ESC to abort measuring.
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Check drying time switch-off
criterion or ∆M
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Switch equipment off and on
again. If the error message
remains inform manufacturer.
Switch equipment off and on
again. If the error message
remains inform manufacturer.