5 How To Replace The Halogen Lamp

1).Remove the stainless steel plate, unscrew four screws.
2).Please note the connection position, 1 connect the mainboard, 3 connect the AC power
supplier,2 connect with the temperature sensor. And please give some mark when you move
the joint. If make mistake will broken the mainboard.So this step is most important.
3).Unscrew the 8 screws on the baffle
4).Unscrew the 2 screws, so we can remove the lamp.