13.1.4 Apportioning mode

Use this function if you wish to increase display speed,
e.g. during apportioning. However, please note that the balance is very susceptible to
ambience conditions. You can set the degree of sensitivity according to site (qui-
et/unquiet).
Repeatedly press the [CAL] key until "PoUr inG" appears.
P0VR�1nG�
Press the [TARE] key; display shows current setting of sensitivi-
nQlnAL.Eno�
ty
Setting of this mode can be confirmed only in settings check
display (see ch. 11.6).
You can use the [CAL] key to select between the following set-
5TABL.Eno�
tings:
"StAbl.Env"
"normL.Env"
nQlnAL.Eno�
" UnStAbl.Env " very unquiet site/insensitive but slow
The current setting is indicated by the standstill display ( ).
Un5TABL.Eno�
Confirm your selection by pressing the [TARE] key
For mode setting in setting control display (see ch.12.6):
very quiet site/sensitive and fast
PQ.S.B1T�
normal site/ medium setting
PQ.n.B1T�
very unquiet site/insensitive but slow
PQ.U.B1T�
48
very quiet site/sensitive and fast
normal site/ medium setting
ABT-BA-e-1312