8.4 Carrying out alternative adjustment methods

Here you start adjustment by selecting a setting in the menu.

8.4.1 Adjustment with internal weight

Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press the [TARE] key. "CAL" is displayed.
Press the [TARE] key. "E CAL" is displayed.
Repeatedly press the [CAL] key until "i CAL" appears.
Press the [TARE] key. The display changes automatically in this
order: "CAL 2", "CAL 1", "CAL 0", and "CAL End".
After successful adjustment the balance automatically returns to
weighing mode.
In case of an adjustment error (e.g. objects on the weighing plate)
the display will show an error message, repeat adjustment.
ABT-BA-e-1312