Density Determination; Density Determination Of Solids - KERN ABT series Operating Instruction

Analytical balance
Hide thumbs Also See for ABT series:
Table of Contents

Advertisement

14.3 Density determination

Density determination with the help of the flush-mounted platform facility is described
in the following.
Density determination becomes even easier when an optional set for density deter-
mination is applied. For further information please refer to the operating instructions
enclosed with the set for density determination.
1. Remove the below-weigh hook cover on the floor of the balance, after remov-
ing the two fixing screws.
2. Hang the user-prepared hanging pan from the below-weigh hook and sink that
hanging pan into the sample liquid in the tank.

14.3.1 Density determination of solids

Condition: Function ",d" (density of solids) is activated. See Chapter 11.3.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly actuate the [CAL] key until "LSG SEt" appears
Press the [TARE] key. The currently set density for the liquid to be
measured appears. In the upper part of the display panel, the
(Example)
ical input status. The leftmost digit blinks.
Enter density for your liquid to be measured. When the [UNIT] key
is pressed, the numerical of the blinking digit increases by 1 at a
(Example)
time. You can determine the value of the flashing digit, or shift the
flashing digit by one position to the right, by pressing the [PRINT]
key. Confirm your setting by pressing the [TARE] key.
Repeatedly press the [ON/OFF] key until the balance is in weigh-
ing mode.
56
symbol and the # symbol appear in order to indicate numer-
ABT-BA-e-1312

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Abt-ba-e-1312Abt 100-5mAbt 120-4m

Table of Contents