11.5 Underfloor weighing

Objects unsuitable for placing on the weighing pan due to size or shape may be
weighed with the help of the flush-mounted platform.
Proceed as follows:
Switch off balance.
Open the closing lid (1) on the bottom of your balance.
• Place your balance over an opening.
• Suspend the goods to be weighed from the hook and carry out the weighing.
(1)
Fig. 1: Underfloor weighing device
• Always ensure that all suspended objects are stable enough to hold the desired
goods to be weighed safely (danger of breaking).
• Never suspend loads that exceed the stated maximum load (max) (danger of
breaking)
Always ensure that there are no persons, animals or objects that might be dam-
aged underneath the load.
After completing the underfloor weighing the opening on the bottom of the balance
must always be closed (dust protection).