12.0 Treatment Report Test

Fluke Impulse 7000 DP Analyzer (or equivalent)
Equipment
Do this...
12.1 Connect the multifunction cable (MFC) to the
Analyzer, and turn on the Analyzer.
12.2 Turn on the Propaq MD unit, and press the
following quick access keys:
a)
b)
c)
12.3 Select Yes.
12.4 Press the CHARGE button, select 200 joules,
then press the CHARGE button again.
12.5 Press the Shock button.
12.6 Wait 20 seconds, press the Treatment quick
access key(
marker.
12.7 Wait 15 seconds and turn off the unit, then wait
2 minutes and turn on the unit.
12.8 On the Propaq MD unit, press the following
quick access keys:
a)
b)
c)
12.9 Highlight the second event and press Select.
12.10 Select Print Treatment Summary.
12.11 Turn off the Propaq MD unit.
9650-0801-01 Rev. B
LOG
R
), and select the first code
LOG
Propag MD Service Manual
Propaq MD Overview
Observe this...
The Clear Log window
displays.
The unit displays a clearing
the log message.
The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.
The unit discharges the shock.
The Treatment Summary
window displays.
Verify that a green check mark
displays next to the file.
Verify that the recorder prints
out the defib event and the
code marker snapshot.
Pass/Fail
o o
o o
o o
o o
o o
o o