Removing The Front Enclosure Module - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Removing the Front Enclosure Module

Tools Required
#2 Phillips screwdriver with torque capability
Orange (wooden) stick
Setup
This procedure should be conducted on a static mat.
Wear a wrist strap to ground during the procedure.
Complete these procedures in the following order before performing this procedure:
"Removing the Handle" on page 57
1.
2.
"Disassembling the Main Unit" on page 60
To reinstall the front enclosure module, reverse the steps.
Front Enclosure Disassembly
1
Remove tape from the top of the unit.
2
Separate monitor front enclosure.
3
Pivot top up with cables still connected.
4
Remove grounding screw as shown in the
illustration.
Note: During reassembly, torque screw to 6
in-lbs.
5
Disconnect the front panel multi-wire cable
in the area illustrated.
9650-0801-01 Rev B
ZOLL Propaq MD Service Manual
Removing the Front Enclosure Module
grounding
screw
multi-wire
cable
63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents