Table 3.1 Error Message Chart
Error Number
1 through 9,
10 through 19,
20 through 29
30 through 39
40 through 45
46 through 49
50 through 59
60 through 69
70 through 79
B30162AA
Problem
System error
After rotor comes to a complete
stop, turn the power off (O),
then back on (I) to reset.
Speedometer error
After rotor comes to a complete
stop, turn the power off (O),
then back on (I) to reset.
Motor error
Turn the power off (O). Ensure
ventilation.
EEPROM error
After rotor comes to a complete
stop, turn the power off (O),
then back on (I) to reset.
Temperature error
• Allow to slow down
(applies only to
• Power off
Microfuge 20R
• Allow to cool down
refrigerated
• Provide better ventilation
centrifuges)
Imbalance error
• Allow to slow down
• Power off
• Allow to cool down
Door error
• Press door key
• Close door
• Remove foreign matter
Process error
• Allow to slow down
• Cycle power
Communication error
• Allow to slow down
• Cycle power
Recommended Action
from the opening of the
door lock device
3
Troubleshooting
User Messages
Explanation
All of these errors stop
the centrifuge or
cause it to decelerate
brakeless.
Errors 34, 35 and 36
will cause the
centrifuge to stop;
Errors 37 and 38 will
not cause the
centrifuge to stop,
these errors will only
return error messages.
Errors 50 and 51 will
cause the centrifuge
to stop.
"With error 60
message "power
failure during run",
with error 61, the
message "stop after
power on"
3-3