Installing the Instrument
• Do not subject the centrifuge to thermal stress. For example, do not position the centrifuge
near a source of heat.
• Do not position the centrifuge where it receives direct sunlight (UV radiation).
• The table or bench that the centrifuge is installed on must be stable and have a solid, flat
surface.
• After moving the centrifuge from cold to warmer rooms, condensation may collect inside the
centrifuge. It's important to allow at least 24 hours for drying before using the centrifuge again.

To install the centrifuge

1
The centrifuge ships in a cardboard box. For easy access, remove the top of the box, the foam
insert on top of the centrifuge, and then the upper part (sides) of the box and set them aside.
a. Then, with the help of another person, move the centrifuge to its final position.
• (Note the warning above regarding centrifuge weight.)
2
Position the centrifuge on a level surface, such as a sturdy table or laboratory bench that is able
to support the weight of the centrifuge and resist vibration.
• Refer to
Specifications
a. Make sure that the centrifuge front feet are fully supported on the table.
b. Locate the centrifuge in an area with sufficient ventilation to allow for heat dissipation.
c. Check that there are 30 cm (12 in.) clearances at the sides of the centrifuge to ensure
sufficient air circulation.
Dimensions are shown in
The centrifuge must have adequate air ventilation to ensure compliance to local
requirements for vapors produced during operation.
For ambient temperature ranges see
Microfuge 20 Nonrefrigerated Model
NOTE
A-2
Microfuge 20 Nonrefrigerated Model Specifications
in
CHAPTER 1
for weight.
Figure A.1
During transport between areas with varying temperatures, condensation may occur inside
the centrifuge. Allow sufficient drying time before running the centrifuge.
or
Microfuge 20R Refrigerated Model
or
Figure
A.2.
Microfuge 20R Refrigerated Model Specifications
Specifications.