The display with show either
A750
=
x.xxx
Std #1 Rep #2
>=Continue
•
Press the right arrow key to return to the assay.
After the last replicate of the last standard is measured, SmartSpec 3000 will construct the
standard curve.
C. Editing standard curve data. After the absorbance of the last replicate of the last
standard has been read, SmartSpec 3000 will offer you a chance to edit the data set before
constructing the standard curve. You can re-read a standard or any replicate of a
standard group, or you can edit the concentration information for any standard or standard
replicate group.
Do you wish to modify
the Standard Curve? No
•
If you do not want to edit the data, press Enter and skip to D below. If you need to
edit the data or to re-read one or more samples, press Select to toggle from No to Yes
and press Enter.
1. Edit Concentrations
2. Reread a Std
•
To edit a concentration, press 1.
Select Standard to edit:
Standard
•
Press Select until the standard you want to edit is displayed and then press Enter.
Change Conc Std #1 from
123.45678 to
•
Use the number keys to input the correct concentration and press Enter.
1. Edit Concentrations
2. Reread a Std
•
Press the left arrow key to return to the assay and complete construction of the
standard curve. If you want to re-read a standard, press 2.
Enter Standard#
Enter Rep#
•
Use the number keys to input the number of the standard that you want to re-read,
press Enter, input the replicated number and press Enter again.
Do you want to reread
blanks each time?
•
You can choose to blank the instrument again before you re-read your sample by
pressing Enter at this screen. If you do not want to re-zero SmartSpec 3000, press
Select to toggle from Yes to No and press Enter.
A750
=
x.xxx
or
Std #1 Mean
<=Back
#1:
123.4567
<=Back
Yes
37
+-
x.xxx
>=Continue