10 Supplement
Ambient conditions
Ambient, storage and transport temperature
Process conditions
Process pressure
Process temperature (thread or flange temperature)
FKM (Viton)
-
EPDM
-
FFKM (Kalrez 6375)
-
1
80 °C
(176 °F)
60 °C
(140 °F)
40 °C
(104 °F)
20 °C
(58 °F)
-40 °C
0 °C
(-40 °F)
(32 °F)
Fig. 30: Ambient temperature - Process temperature
1
Ambient temperature
2
Process temperature (depending on the seal material)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
4)
Single chamber housing
-
52
-40 ... +80 °C (-40 ... +176 °F)
-1 ... +40 bar/-100 ... +4000 kPa (-
-14.5 ... +580 psig), depending on the process
fitting
-40 ... +150 °C (-40 ... +302 °F)
-40 ... +150 °C (-40 ... +302 °F)
-20 ... +150 °C (-4 ... +302 °F)
The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.
50 °C
100 °C
(122 °F)
(212 °F)
l
or:
l
or:
l
or:
l
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
4)
8" FF.
150 °C
2
(302 °F)
1 x cable gland M20 x 1.5 (cable: ø 5 ... 9 mm),
1 x blind stopper M20 x 1.5
1 x closing cap M20 x 1.5; 1 x blind stopper
M20 x 1.5
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5