10 Supplement
Installation reference conditions
Min. distance to installations
-
Vessel
-
Medium
-
Installation
-
Sensor parameter adjustment
Deviation
48
> 500 mm (19.69 in)
metallic, ø 1 m (3.281 ft), centric installation,
process fitting flush with the vessel ceiling
Water/Oil (Dk value ~2.0)
Probe end does not touch the vessel bottom
Gating out of false signals carried out
In bulk solids applications, implementing a false
signal suppression is generally not recommended.
The accuracy of bulk solid applications depends
considerably on the application. Hard and fast
accuracy specifications are thus not available.
see diagrams
Depending on the installation conditions, there can
be deviations which can be rectified with an
adaptation of the adjustment or a change of the
measured value offset in the DTM service mode.