4 Mounting
Inflowing medium
Fixing
16
Make sure that the probe is not subjected to strong lateral forces.
Mount VEGAFLEX 61 at a position in the vessel where no
disturbances, e.g. from filling openings, agitators, etc., can occur.
Fig. 9: Lateral load
If there is a danger of the probe touching the vessel wall during
operation due to product movements or agitators etc., the measuring
probe should be securely fixed.
In the gravity weight there is a thread (M12), e.g. for a ring bolt
(optional) - (article no. 2.27424).
Make sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.
Avoid undefined cable-vessel connections, i.e. the connection must be
either grounded reliably or isolated reliably. Any deviation from this
requirement can lead to measurement errors.