GLUCOSE READINGS

Treatment decisions should not be based solely on results from
the SEVEN PLUS. You must confirm with a blood glucose meter
before making therapeutic adjustments.
You must update SEVEN PLUS calibration every 12 hours at a
minimum. The performance of the System when calibrated less
frequently than every 12 hours has not been studied.
Press any button on the Receiver and the following glucose information will be shown to you:
68
|
SEVEN PLUS Continuous Glucose Monitoring System
Low Glucose
Alert Setting = 80 mg/dL (Default)
High Glucose
Alert Setting = 200 mg/dL (Default)
Most recent 5-min-
ute glucose reading
(160 mg/dL) Status
Box
Trend Arrow
Time (A=AM, P=PM)
Trend Graph Hours
Plot of the most
recent glucose reading
(160mg/dL) on the 3 hour
Glucose Trend Screen