10
Guidance and Manufacturer's Declaration- Electromagnetic Immunity
The Receiver is intended for use in the electromagnetic environment specified below. The
customer or the user of the Receiver should assure that it is used in such an environment.
Immunity
IEC 60601
Test
Test Level
Conducted
3 Vrms
RF
150 k Hz to
IEC 61000-
80 MHz
4-6
Radiated RF
3 V/m
IEC 61000-
80 MHz to
4-3
2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people
128
|
SEVEN PLUS Continuous Glucose Monitoring System
Compliance
Electromagnetic Environment Guidance
Level
Portable and mobile RF communications equipment
should be used no closer to any part of the Receiver,
including cables than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance
d =1.2 P
1/2
d = 1.2 P
1/2
d = 2.3 P
1/2
Where P is the maximum output power rating of the
3 V
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
be less than the compliance level in each frequency
3 V/m
range
.
b
Interference may occur in the vicinity of equipment
marked with
following symbol:
80 MHz to 800 MHz
800 MHz to 2.5 GHz
should
a