Automatic Measurement / Traffic; Automatic Measurement / Airline - Dräger Alcotest 6510 Instructions For Use Manual

Breath alcohol screening device
Hide thumbs Also See for Alcotest 6510:
Table of Contents

Advertisement

Automatic measurement / Traffic
— The mode "Traffic" was set by the manufacturer:
ZERO: <4 µg/100 mL
PASS: <31 µg/100 mL
ALERT: <36 µg/100 mL
FAIL: ≥36 µg/100 mL
Set mode see page 18.
— The sampling is activated automatically after the minimum
breathing volume and minimum blowing period are reached.
Preparation
Insert new mouthpiece in the holder
(page 10).
Switch instrument on (page 11).
After 2 seconds:
— the user info screen
or
"WAIT" is indicated on the display.
The user info screen can also be viewed
by pressing the
- button. Close with
the
- button
OK
After approximately 6 seconds "READY" is
displayed and a short audible signal is given.
At the same time the current test number is
indicated. A green light flashes.
The instrument is ready for measurement.

Automatic measurement / Airline

— ZERO: <4 µg/100 mL
PASS: <9 µg/100 mL
FAIL: >9 µg/100 mL
Set mode see page 18.
— The sampling is activated automatically after the minimum
breathing volume and minimum blowing period are reached.
Preparation
Insert new mouthpiece in the holder
(page 10).
Alcotest 6510
Switch instrument on (page 11).
After 2 seconds:
— the user info screen
or
WAIT
"WAIT" is indicated on the display.
The user info screen can also be viewed
by pressing the
the
After approximately 6 seconds "R.AIR" is dis-
played and a short audible signal is given. At
the same time the current test number is indi-
READY
cated. A green light flashes.
Test No. 00140
The instrument is ready for measurement.
- button. Close with
- button
OK
Alcotest 6510
WAIT
R.AIR
Test No. 00140
13

Hide quick links: