Technical data

Measuring principle
Electrochemical sensor
Measurement range
Breath alcohol concentra-
0 to 250 µg/100mL
tion
(Ethanol-Mass per breath volume at
34
Display range
Display range is identical to measuring range.
If the result is exceeding the measuring range the display will
show three arrows "↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ " as overflow indication.
Ambient conditions
For operation
0 to 40
10 to 100 % r.h.
(non condensing)
600 to 1400 hPa
For storage
–40 to 65
Sample conditions
minimal blowing time
2 s default
minimal volume
1.2 L default
24
o
C and1013 hPa.)
o
o
C (32 to 104
F)
o
o
C (–40 to 149
F)
Measurement accuracy
Reproducibility with an ethanol standard
up to 50 µg/100 mL
0.8 µg/100 mL or
≥ 50 µg/100 mL
1.7 % of measurement value
the higher value is valid
up to 1.00 ‰
0.017 ‰ or
≥ 1.00 ‰
1.7 % of measurement value
the higher value is valid
Sensitivity drift
typically 0.6 % of meas. value/month
Intervals before readiness for operation
1st measurement
about 6 s
after terminating a result of a previous test in range:
(breath alcohol unit)
25 µg/100 mL
120 s
50 µg/100 mL
140 s
100 µg/100 mL
180 s
> 150 µg/100 mL
120 s
Dimensions
140 mm x 80 mm x 35 mm (W x H x D)
Weight
approx. 195 g
Electrical supply
Batteries
2 x 1.5 V Type Alkaline
(Mignon, LR6, AA)
approx. 1500 measurements
CE-marking
electromagnetic compatibility
(Directive 89/336/EEC)