System Accuracy Results for Glucose
Concentrations ≥5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5%
Within ±10%
Within ±15%
62.9% (283/450)
94.7% (426/450)
99.3% (447/450)
System Accuracy Results for Glucose
Concentrations between 2.1 mmol/L
and 25.9 mmol/L
Within ±0.83 mmol/L or ±15%
99.5% (597/600)
NOTE: Where 2.1 mmol/L represents the lowest glucose
reference value and 25.9 mmol/L represents the highest
glucose reference value (YSI value).