System Accuracy - OneTouch Verio Reflect Owner's Booklet

Hide thumbs Also See for Verio Reflect:
Table of Contents

Advertisement

Available languages

Available languages

12
System information

System accuracy

Diabetes experts have suggested that 95% of glucose
meter results should agree within 0.83 mmol/L of a
laboratory method when the glucose concentration is
lower than 5.55 mmol/L, and within 15% of a laboratory
method when the glucose concentration is 5.55 mmol/L
or higher.
Samples from 100 patients were tested using both
the OneTouch Verio Reflect
System and the YSI 2900
Glucose Analyzer laboratory instrument.
System accuracy results for glucose
concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
Within
Within
±0.28 mmol/L
±0.56 mmol/L
±0.83 mmol/L
73.7% (137/186)
96.8% (180/186)
100% (186/186)
158

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents