System Accuracy - OneTouch VERIO Owner's Booklet

Table of Contents

Advertisement

8
Detailed information about
your system

System Accuracy

Diabetes experts have suggested that glucose meters
should agree within 15 mg/dL of a laboratory method
when the glucose concentration is lower than 75 mg/dL,
and within 20% of a laboratory method when the glucose
concentration is ≥75 mg/dL or higher. Samples from 100
patients were tested using both the OneTouch® Verio®
System and the YSI 2300 Glucose Analyzer laboratory
instrument.
System Accuracy Results for Glucose
Concentrations <75 mg/dL
Percent (and number) of meter results that match the
laboratory test
Within
±5 mg/dL
50.7%
(70/138)
Within
±10 mg/dL
92.0%
(127/138)
116
Within
±15 mg/dL
100%
(138/138)