Detailed information about your system

System Accuracy

Diabetes experts have suggested that glucose meters
should agree within 0.83 mmol/L of a laboratory
method when the glucose concentration is lower than
5.55 mmol/L, and within 15% of a laboratory method when
the glucose concentration is 5.55 mmol/L or higher.
Samples from 100 patients were tested using both the
OneTouch Verio Flex™ System and the YSI 2300 Glucose
Analyser laboratory instrument.
System Accuracy Results for Glucose
Concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28 mmol/L
82.0% (123/150)
Within
±0.56 mmol/L
98.0% (147/150)
101
Within
±0.83 mmol/L
100% (150/150)
9