System Accuracy

Diabetes experts have suggested that glucose meters
should agree within 15 mg/dL of a laboratory method
when the glucose concentration is lower than 75 mg/dL,
and within 20% of a laboratory method when the glucose
concentration is ≥75 mg/dL or higher. Samples from 100
patients were tested using both the OneTouch® Verio®
System and the YSI 2300 Glucose Analyzer laboratory
instrument.
System Accuracy Results for Glucose
Concentrations <75 mg/dL
Percent (and number) of meter results that match the
laboratory test
Within
±5 mg/dL
50.7%
(35/69)
Within
±10 mg/dL
89.9%
(62/69)
114
Within
±15 mg/dL
100.0%
(69/69)