System Accuracy

Diabetes experts have suggested that glucose meters
should agree within 0.83 mmol/L of a laboratory method
when the glucose concentration is lower than 4.2 mmol/L,
and within 20% of a laboratory method when the glucose
concentration is ≥4.2 mmol/L or higher. Samples from
100 patients were tested using both the OneTouch® Verio®
System and the YSI 2300 Glucose Analyzer laboratory
instrument.