Yellow= 5mm Red=7mm margin
Detailed information about your system
9
System Accuracy Results for Glucose
Concentrations *5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5%
63.3%
(266/420)
System Accuracy Results for Glucose
Concentrations between 1.5 mmol/L
and 26.8 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±0.83 mmol/L or ±15%
NOTE: Where 1.5 mmol/L represents the lowest glucose
reference value and 26.8 mmol/L represents the highest
glucose reference value (YSI value).
Type: Project File Name: 3130451 Revision: 1
Effectivity Date: Mar 18 2015
Statement of Use: Verify status before each use
PF3130451Rev1_OTSPF_OB_I_GB_en_zug_R2.indd 106
Within ±10%
Within ±15%
95.5%
99.8%
(401/420)
(419/420)
98.8% (593/600)
106
Status: Current
State: Released
3/12/15 11:36 AM