Symbols that appear on the display
Symbols
Appears while measurement is in progress.
It blinks when the pulse is detected.
Irregular Heartbeat symbol (I.H.B.)
Appears when an irregular heartbeat
is detected.
It may light when a very slight vibration
like shivering or shaking is detected.
Previous measurements stored in memory.
Illuminates in order from bottom
when the
add
constant speed exhaustion at the
auscultation mode
Illuminates when the auscultation
mode is ON.
FULL BATTERY
The battery power indicator during
measurement.
LOW BATTERY
The battery power is low when it
blinks.
Illuminates when the AC adapter is
connected to the device.
Blinks while the battery is being
recharged.
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL DISPLAY ERROR
The pulse is not detected correctly.
Function / Meaning
button is pressed to
the
pressurization
English 8
Recommended Action
Measurement is in progress.
Remain as still as possible.
during
Recharge the device
using the AC adapter.
Take another measurement.
Remain still during
measurement.
Apply the cuff correctly,
and take another
measurement.