RS-232 CABLE CONFIGURATIONS
Appendix N
RS-232 Cable configurations
This appendix supplies additional information about the RS-232 cable
configurations between the oscilloscope and a connected device. The
oscilloscope and most of the devices are Data Terminal Equipment (DTE)
configurated.
Communication lines:
TxD
= Transmitted Data
RxD
= Received Data
RTS
= Request To Send
CTS
= Clear To Send
DTR
= Data Terminal Ready
DSR
= Data Set Ready (= Modem Ready)
DCD
= Data Carrier Detect
S.GND = Signal ground
F.GND = Safety ground
A. Cable to printer/plotter/controller with software handshake
parameters: "3-wire/7-wire" = 3-wire
ATTENTION:
The maximum Baud rate for the PM8277 and PM8278 plotters is
4800 baud. For a higher Baud rate you must use a hardware
handshake configuration.
NOTE:
The LQ1500 printer can have a 6-pole DIN connector with the
following connections:
TxD on pin 1
RxD on pin 3
"XON-XOFF on off" = on
OSCILLOSCOPE
(DTE)
SIGNAL
9 PIN
(female)
TxD
3
RxD
2
RfR
7
CTS
8
DTR
4
DSR
6
1
DCD
S.GND
5
CASE
F.GND
DEVICE
(DTE)
9 PIN
3
2
7
8
4
6
1
5
CASE
DTR on pin 2
S.GND on pin 5
25 PIN
2
3
4
5
20
6
8
7
1
ST6930
N - 1