FUNCTION REFERENCE
NO DTB, ACQ. TOO LONG
NO DTB IN ROLL MODE
NO ENVELOPE IN ROLL MODE
NO EVENT DELAY, DUAL SLOPE
NO EVENT DELAY IN TV MODE
NO IND WITH THIS COUPLING
NOT PART OF X VS Y REGISTER
PLEASE FIRST SWITCH TO DSO
PLOT KEY = ABORT
PRESS 2 SEC FOR AUTOCAL
PROBE DETECTED, NO CHANGE
Del'd TB only possible at acquisition
length of 512 sample points.
Functions 'ACQUISITION LENGTH' and
'DEL'D TB'.
Del'd TB and Roll modes cannot be
combined.
Function 'DEL'D TB' and 'TB MODE'.
Envelope and Roll modes cannot be
combined.
Function 'ENVELOPE' and 'TB MODE'.
Event delay and dual slope cannot be
combined.
Function 'DELAY'.
Event delay and TV trigger mode cannot
be combined.
Function 'DELAY' and 'TV TRIGGER'.
Trigger level indication not possible at dc
input coupling and ac or lf-rej trigger
coupling.
Function 'UTILITY SCREEN & SOUND.
X-source only possible out of chosen
register.
Functions 'DISPLAY MENU' and
'X-DEFLECTION'.
Activated function only possible in digital
mode.
Function 'ANALOG/DIGITAL MODE'.
Indicates that the print/plot action can
only be aborted with PLOT key.
Function 'PRINTING AND PLOTTING'.
CAL key must be pressed for more then
two sec to start autocalibration.
Function 'CALIBRATION AUTOCAL'.
The automatic probe detection overrules
manual selection when a probe with
indication ring is used.
Function 'PROBE UTILITIES'.
5 - 65