Fluke CombiScope PM3370B User Manual page 91

Autoranging combiscope instrument
Hide thumbs Also See for CombiScope PM3370B:
Table of Contents

Advertisement

4 - 46
CURSORS
-
Use the softkeys, to select the amplitude cursors ('=' intensified).
Two amplitude cursors will appear in the display.
The menu permits the channel to be selected for which the amplitude cursors
apply.
When channel 1 is the only channel displayed, 'ch1' is the only selection.
-
Turn the TRACK
control to move both
cursors, and turn the
Delta ( ) control to
move only the
cursor.
CURSORS
Amplitude and Time cursors can be displayed at the same time. When both
amplitude and time cursors are displayed, you must select which cursors are
affected by the TRACK and
softkey 'CONTROL'.
Channel selection applies only to the voltage cursors. The time cursor always
applies to all channels.
To use both types of cursors at the same time, do the following:
-
Press the softkey next to '= || #' so that '#' is intensified.
-
Press the CONTROL softkey so that '=' is intensified.
-
Turn the TRACK control and observe that only the amplitude cursors
move.
-
Press the CONTROL softkey so that '||' is intensified.
-
Turn the TRACK control and observe that only the time cursors move.
HOW TO USE THE INSTRUMENT
s V=360mV
ch1: T=
TRACK
CH1
200mV
controls. This selection is made by toggling the
MTB 200us
ch1
AMPLITUDE & TIME
VOLT
CURSORS
on off
#
-
ch1
-
READOUT
ST6731
9303