Signal Wiring Recommendations; Testing The System - Teledyne 2010A Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer

2-11 Signal Wiring Recommendations

The signal leads are the bias and sensor leads ref fig. 3.4, and reference
D-70731 interconnection diagram. The maximum interconnection distances
between the 2010 control unit and the analysis unit is limited by the sensor
signal requirements. The sensor has two limiting factors. The first is series
resistance which is limited to a maximum of 2.5 ohms. Refer to the intercon-
nection diagram for wire gages and distances which will meet this require-
ment.
The second issue is one of signal quality. At greater distances prob-
lems associated with signal degradation increases. Signal degradation prob-
lems have two major issues. The first is associated with the local installation
environment. The number, proximity, and power associated with local
sources of EMI radiation can not be accounted for by TAI. These issues can
only be controlled by the end user. The other factor is the amplitude of the
signals generated by the thermal conductivity cell. This signal level is a
function of the gas composition of the gases involved. This signal level can
vary from micro-volts to mili-volts dependent on application. Consult TAI
for recommendations based on sample stream composition. Signal quality
and operating distances can be increased by the uses of better quality
shielded low loss cable. The cable configuration should be twisted pair with
double shielding for best performance. The signal cable should not be run
with AC power or digital signals.

2-12 Testing the System

After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete,
Figure 3-6: FET Series Resistance
Teledyne Analytical Instruments
Part II: Analysis Unit
Part II: 2-7

Advertisement

Table of Contents
loading

Table of Contents