3.3.9 Remote Probe Connector

The 2010A is a single-split configuration analyzer, the Remote Probe
connector is used to interface the Analysis Unit. See Figure 3-6.
Figure 3-: Remote Probe Connector Pinouts
The voltage from these outputs is nominally 0 V for the OFF and
15 V dc for the ON conditions. The maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
3-0 Part I
Table 3-4: Required Rs-232 options
Parameter
Baud
Byte
Parity
Stop Bits
Message Interval
Teledyne Analytical Instruments
Setting
2400
8 bits
none
1
2 seconds
Model 2010A