4 Operation
When the testing is complete, the results are displayed.
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The results
screen alternates for a time with:
Then the analyzer returns to the initial System screen.

4.3.6 The Model Screen

Move the < > arrow key to MORE and press Enter. With MODEL
blinking, press Enter. The screen displays the manufacturer, model, and soft-
ware version information.

4.3.7 Checking Linearity with ALGORITHM

From the System Function screen, select ALGORITHM, and press
Enter.
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.
Press Enter again.
Select and Enter VERIFY to check whether the linearization has been
accomplished satisfactorily.
4-10 Part I
RUNNING DIAGNOSTIC
Testing Preamp — Cell
Power: OK Analog: OK
Cell:
2 Preamp: 3
Press Any Key
To Continue...
sel rng to set algo:
—> Ø1 Ø2 Ø3
Gas Use: H2 — N2
Range:
Ø — 10%
Algorithm setup:
VERIFY SET UP
Dpt INPUT
OUTPUT
Ø Ø.ØØ
Ø.ØØ